[MRI] (영/한) 3D TOF MRA / 3D 자기공명 혈관 조영술
(English)
3D TOF MRA?
This is a method of acquiring a 3D image by performing z-axis encoding in the slice selection direction in addition to the frequency direction and the phase direction.
Like 2D TOF MRA, it is a method that uses the flow related signal enhancement phenomenon of the inflowing blood flow and is the most used method to view the vessels of the head.
Compared to 2D TOF MRA, SNR is better and images of very thin slices are possible, reducing intravoxel dephase by reducing the voxel size. Because it is less affected by turbulence, there is less signal loss or imagination in the stenosis region, and it can be used for diagnosis of vascular lesions in the cranial cavity because images with high spatial resolution are possible.
The blood flow rate should be fast enough for the blood flow to pass all image slabs between successive RFs, so that the effect of inflow is sufficient and good blood flow signal strength is exhibited. Therefore, it is more sensitive to the saturation effect and has a disadvantage that the slow blood flow cannot be captured well.
It is useful for observing cerebral arteries with high blood flow speed and is used for diagnosis of cerebral artery malformations (AVM) and aneurysms.
The magnetization transfer (MT) technique used in the TOF technique is a method used to attenuate the background signal to increase tissue contrast.
By applying off-resonance RF pulses to the gradient echo sequence, thin blood vessels appear to be applied to the TOF, which will reduce the signal intensity of muscle or brain gray matter without changing the signals of fat, cerebrospinal fluid (CSF), and blood.
The off-resonance technique is a method of saturating only the non-fluid protons of a macromolecule without including the signal of the water molecule by using a chemical shift selective (CHESS) pulse whose center frequency is a frequency several kHz away from the RF of the water molecule. To this, a fat suppression technique is added to attenuate the bright signal intensity from the fat.
Disadvantages, as in the 2D TOF, the short T1 material shows high signal intensity, so subacute bleeding methemoglobin in the cranial cavity appears to have a high signal intensity similar to blood flow or superimposed on the signal of the blood flow. Therefore, when you scan, it is necessary to determine the direction of the image slab so that the blood flow is not saturated in consideration of the blood flow at the scan range.

by radiographer nara
(한국어)
3D TOF MRA?
주파수 방향과 위상 방향 이외에 절편 선택 방향으로 z축 부호화를 해줌으로써 3D 이미지를 획득하는 방식이다.
2D TOF MRA와 마찬가지로 유입되는 혈류의 유속신호강조현상을 이용하는 방법이며 두부의 혈관을 보기 위해 가장 많이 사용하는 방법이다.
2D TOF MRA에 비해 SNR이 좋고 매우 얇은 절편의 영상이 가능하여 복셀 크기를 감소시켜 intravoxel 탈위상을 줄일 수 있다. 난류에 영향을 덜 받으므로 협착 부위의 신호 손실 또는 허상이 덜하며 높은 공간분해능의 영상이 가능해 두개강 내 혈관성 병변의 진단에 많이 이용된다.
혈류 속도는 혈류가 연속적인 RF 사이에 이미지 슬랩을 모두 지날 수 있을 정도로 빨라야 유입되는 효과가 충분히 나타나 좋은 혈류 신호 강도가 나타나게 된다. 그러므로 포화효과에 더 민감하고 느린 혈류는 잘 포착이 안되는 단점이 있다.
혈류 속도가 빠른 뇌동맥의 관찰에 유용하고 뇌동맥기형, 동맥류 등의 진단에 사용된다.
TOF 기법에 사용되는 자화전이 기법은 배경 신호를 감쇄시켜서 조직의 대조도를 증가시키는데 쓰는 방법이다.
그라디언트 에코 시퀀스에 off-resonance RF 펄스를 줌으로써 지방, 뇌척수액, 혈액 등의 신호에 변화를 주지 않고 근육이나 뇌의 회백질의 신호강도를 감소 시킬 TOF에 적용함으로 가는 혈관이 잘 나타난다.
off-resonance 기법은 물분자의 공명주파수로부터 수 kHz 떨어진 주파수를 중심주파수로 하는 CHESS(chemical shift selective) 펄스를 사용하여 물분자의 신호는 포함시키지 않으면서 거대분자의 비유동성 양성자들만 포화시키는 방법이다. 여기에 지방소거법을 추가하여 지방에서 나오는 밝은 신호강도를 감쇄시킨다.
단점으로는 2D TOF에서와 마찬가지로 짧은 T1 물질이 고신호 강도로 보여 두개강 내 아급성 출혈 메트 헤모글로빈이 혈류와 유사한 고신호 강도로 보이거나 혈류의 신호에 겹쳐서 보인다. 그러므로 검사하고자할 때는 검사부위의 혈류를 고려하여 혈류가 포화되지 않도록 이미지 슬랩의 방향을 결정할 필요가 있다.