방사선사나라 Radiographer Nara

[MRI](영/한) FSE (Fast Spin Echo), TSE(Turbo Spin Echo) / 고속 스핀에코 본문

자기공명영상 (MRI)

[MRI](영/한) FSE (Fast Spin Echo), TSE(Turbo Spin Echo) / 고속 스핀에코

SEONARA 2020. 4. 24. 12:15
반응형

(영어/영문/English)

 

Classic spin echo (SE) method?

you can get one echo per TR. In the end, you can get an image only by receiving one echo for every TR and filling the K-space with the number of phase encoding steps.

 

Scan time = TR x Phase encoding x NEX
TR affects contrast, Phase encoding affects resolution, and NEX affects SNR.

 

 

Multi echo (ME) method?

Multiple signals are obtained by applying multiple 180º RF pulses in one TR, but the echoes at this time fill only one line of each K-space with the signal applied with the same gradient. So after TR is repeated as many times as the number of phase encoding steps, we can get several kinds of images that have different TE.

 

 

FSE method?

This is a modified version of Spin Ehco's multi echo method for fast imaging. As with the multi echo method, several 180º RF pulses are applied to one TR. This is a method of reducing the scan time by filling only one K-space by applying different size phase encoding gradient to every echo.

Since the FSE image can be obtained in a short time while showing almost the same contrast as the spin echo image, the FSE image of good quality can be made by increasing the NEX or matrix. So FSE image have recently replaced the SE-T2WI.

 

FSE features?
(180º refocusing RF pulse is applied continuously at very short time intervals)

1. The diffusion effect of fat molecules is reduced and the J-coupling (spin-spin coupling) effect of fat molecules is weakened, so that the signal of fat appears brighter than the spin echo image.

2. Since the dephase effect due to the magnetic field inhomogeneity is reduced and the susceptibility effect is reduced, it is less affected by ferromagnetic.
3. The effect of magnetization transfer(MT) occurs well between protons of free water and bound water bound to macromolecules like proteins.

 

ETL (Echo train length) = turbo factor = TSE (Turbo spin echo)?
It refers to several 180º RF pulses applied in one TR. The more ETLs, the shorter the scan time.
However, the larger the number of ETLs, the lower the resolution of the image (T2 blurring) due to the signal attenuation effect by T2 relaxation and is affected by slight motion.
Scan time = TR x Phase encoding x NEX / ETL

 

Effective TE?
It means TE of echo located near the center of K-space, and the contrast of the image is determined by effective TE.
In order to obtain an effective TE, it is necessary to obtain a large signal by applying a low phase encoding gradient.
In order to obtain a T2WI, the long TE echo should be located near the center of the K-space
and to obtain a PDWI, the short TE echo should be positioned near the center of the K-space.

 

 

fast spin echo (FSE) or turbo spin echo (TSE)

by radiographer nara


(국어/국문/Korean)

 

스핀에코 방식?

한번의 TR에 한 개의 에코를 얻는다. 결국 매 TR 마다 하나의 에코를 받아서 위상부호화 수만큼 에코로 K-space를 다 채워야만 영상을 얻을 수 있다.

 

Scan time = TR x Phase encoding x NEX

TR은 대조도, Phase encoding은 해상력, NEX는 SNR에 영향을 준다.

 

 

다중신호 방식?

한 TR안에 여러 개의 180º RF 펄스를 인가하여 여러 개의 신호를 얻지만 이때의 에코들은 동일한 경사자장이 인가된 신호로 각각의 K-space의 한 줄 씩만을 채우므로 위상부호화 수만큼 TR이 반복 된 후 TE가 다른 여러 종류의 영상을 만든다.

 

 

FSE 방식?

스핀에코의 다중신호 방식을 고속 영상법으로 변형한 것으로 다중신호 방식과 마찬가지로 한 TR안에 여러 개의 180º RF 펄스를 인가한다. 이때 발생하는 에코에는 각각 다른 크기의 위상부호화 경사자장을 인가하여 하나의 K-space만을 채우게 함으로 검사시간을 줄이는 방법이다.

스핀에코 영상과 거의 같은 대조도를 나타내면서 빠른 시간에 영상을 얻을 수 있으므로 NEX나 매트릭스를 증가시켜 좋은 화질의 영상을 만들 수 있어 최근에는 스핀에코 T2 강조영상을 거의 대체하고 있다.

 

FSE 특징?
(180º 재초점 RF 펄스를 매우 짧은 시간 간격으로 연속적으로 인가)

1. 지방분자의 확산 효과가 감소하고 지방분자의 J-coupling (spin-spin coupling)효과가 약화되어 지방의 신호가 스핀에코 영상보다 밝게 보인다.

2. 자장의 비균일성에 의한 탈위상효과가 적어지고 자화율 효과가 감소해 강자성에 의한 영향을 적게 받는다.

3. 단백질과 같은 거대 분자와 결합해 있는 결합수와 자유수의 양성자들 사이에서 자화전이 효과가 잘 일어난다.

 

ETL (Echo train length) = turbo factor = TSE (Turbo spin echo)?

한 TR안에 인가되는 여러 개의 180º RF 펄스를 말하며 ETL의 수가 많을 수록 검사시간은 짧아진다.

하지만 ETL의 수가 많을 수록 T2 이완에 의한 신호감쇄효과 때문에 영상의 해상도가 떨어지고 약간의 움직임에도 영향을 받는다.

Scan time = TR x Phase encoding x NEX / ETL

 

유효 에코 시간?

K-space 중앙 부근에 위치한 에코의 TE를 의미하며 이에 따라 영상의 대조도가 결정된다.

유효 에코 시간을 얻기 위해서는 위상부호화 경사자장이 적은 것을 인가하여 큰 신호를 얻어야 한다.

T2 강조영상을 얻기 위해서는 긴 TE의 에코를 K-space 중앙 부근에, 양성자밀도 강조영상을 얻기 위해서는 짧은 TE의 에코를 K-space 중앙 부근에 위치 시켜야 한다.

 