방사선사나라 Radiographer Nara

[MRI] (영/한) Image parameter(6) - FC , Gating / 영상변수(6) - 유동보정, 게이팅 본문

자기공명영상 (MRI)

[MRI] (영/한) Image parameter(6) - FC , Gating / 영상변수(6) - 유동보정, 게이팅

SEONARA 2020. 4. 14. 11:47
반응형

(영어/영문/English)

 

11. FC (Flow Compensation)

Sending a 180º RF pulse to the tissue to be imaged and obtaining an echo signal is the same as the conventional method, and a rephasing gradient magnetic field is added to re-magnetize the phase-shifted proton.

 

The fluid in the spinal cavity vibrates and moves at an amplitude of several centimeters per second, the fastest in the cervical region and slowly slows down to the lumbar region. At this time, the use of flow compensation can minimize artifacts.

 

The pulse waveform that uses the gradient magnetic field of the FC is called 'gradient movement nulling'.
This increases the TE because the gradient magnetic field is used once more in addition to the specified pulse waveform.

If FC is used in the T2 weighted image, the T2 enhancement effect and contrast of the image increase as much as the lengthened TE, but the SNR decreases.
In addition, if the FC is used in the T1 weighted image that collects signals by shortening the TE, the T1 effect of the image decreases as much as the longened TE, so it is not generally used.

 

 

12. Gating

1) Cardiac gating
Although it is not possible to remove completely the motional artifacts caused by the heartbeat, it is a method of synchronizing the movement cycle of the heart to minimize this.

 

When synchronizing, the TR is the same as the cardiac cycle, so the desired TR cannot be applied arbitrarily.
ECG (electrocardiogram) gating technique is used to obtain images of the heart, where the R wave is the starting point and trigger of the heartbeat.
Since the RF pulse is pulsed when the electrical response of the R wave is sensed, the interval at which the RF pulse is given is determined by the patient's heartbeat and can be more easily approached by converting the R-R interval into an effective TR.

 

R-R interval = 60 sec / heart rate per minute

 

R-R interval

2) Respiratory gating
Breathing movements cause blurry images in the imaging of the chest or abdomen. Respiratory gating is a technique used to minimize artifacts caused by breathing.

 

A low gradient magnetic field is used to reduce the effect of motion when the patient inhales, and a steep gradient magnetic field is used at the end of exhalation, the point where movement by the breath is minimal.
Another method of reducing artifacts caused by breathing is the fast image technique, which uses breath holding technique to reduce scan time.

 

3) Peripheral gating
This technique is used to minimize artifacts caused by brain blood flow.
It uses an optical pulse sensor to obtain information about blood pressure flowing through the body by mechanical action.

MRI image parameters by Radiographer Nara 방사선사나라

by radiographer nara


(국어/국문/Korean)

 

11. 유동보정

영상화 하고자 하는 조직에 180º RF 펄스를 보내고 에코 신호를 얻는 것은 기존의 방법과 동일하며, 위상변화된 프로톤을 재자화 시키기 위한 재위상화 경사자장을 부가한다.

 

척수강 내의 수액은 초당 수 cm의 진폭으로 진동하며 움직이는데, 경추 부위에서 가장 빠르고 요추 부위로 이동하면서 서서히 느려진다. 이 때 유동보정을 이용하면 인공물을 최소화할 수 있다.

 

유동보정의 경사자장을 이용하는 펄스 파형을 'gradient movement nulling' 이라고 한다.

이것은 정해진 펄스파형에 추가로 경사자장을 한 번 더 사용하기 때문에 TE가 늘어난다.

T2 강조영상에서 유동보정을 사용하면 길어진 TE 만큼 영상의 T2 강조효과와 대조도가 증가하지만 SNR은 떨어진다.

그리고 TE를 짧게 해서 신호를 수집하는 T1 강조영상에서 유동보정을 사용하면 길어진 TE만큼 영상의 T1효과가 떨어지므로 일반적으로 사용하지 않는다.

 

 

12. 게이팅

1) 심장 게이팅

심박동으로 인한 운동성 인공물을 완전히 제거할 수 없지만, 이것을 최소화 시키기 위하여 심장이 움직이는 주기에 동기화 시키는 방법을 말한다. 

 

동기화 시킬 경우, TR은 심장의 주기와 같기 때문에 임의적으로 원하는 TR을 적용할 수 없다.

심장의 영상을 얻기 위해 ECG 게이팅 기법을 사용하는데, 여기서 R파는 심박동의 출발점이자 트리거라고 할 수 있다.

RF 펄스는 R파의 전기적인 반응이 감지될 때 펄스가 나가게 되므로 RF 펄스가 주어지는 간격인 TR은 환자의 심박동에 의해 결정되고 R-R 인터발을 effective TR로 환산하면 좀 더 쉽게 접근할 수 있다.

 

R-R interval = 60sec/1분당 심박동수

 

2) 호흡 게이팅

호흡에 의한 움직임은 흉부나 복부를 영상화 하는데 있어 흐린 영상을 유발하는데, 호흡 게이팅은 호흡에 의한 인공물을 최소화 하기 위해 사용하는 기법이다.

 

환자의 흡기 때 움직임에 의한 영향을 줄이기 위해 낮은 경사자장을 사용하고, 호흡에 의한 움직임이 최소인 지점인 호기의 말기에서 급경사자장을 사용하게 된다.

호흡에 의한 인공물을 줄이는 또 다른 방법으로 최근 빠르게 이미지를 얻는 기술을 이용하는데, 이는 검사시간을 줄이기 위해 호흡정지기법을 사용한다.

 

3) 말초신경 게이팅

뇌 혈류에 의해서 일어나는 인공물을 최소화 하기 위해 사용하는 기법이다.

이는 광 맥박 센서를 사용하여 신체를 통해 흐르는 혈압을 기계적인 동작으로 정보를 얻는다.

 