방사선사나라 Radiographer Nara

[MRI] (영/한) Nyquist ghost artifact on EPI / 이피아이 기법에서 나이키스트 유령 인공물 본문

자기공명영상 (MRI)

[MRI] (영/한) Nyquist ghost artifact on EPI / 이피아이 기법에서 나이키스트 유령 인공물

SEONARA 2020. 5. 13. 10:41
반응형

(English)

 

The Nyquist ghost artifact on EPI?

When sampling data, the bandwidth must be given at least twice the signal per 1Hz to know the frequency. However, in the case of the EPI pulse sequence, since it has a zig-zag trajectory through k-space, the echo moves slightly from the center point for each phase encoding step.

The artifact that appears due to this is called the Nyquist ghost artifact, and distortion of the image occurs in the phase encoding direction. If there are n pixels in the FOV, this artifact appears to be shifted by n/2 pixels relative to the main image located in the correct position, resulting in making three images.

 

Remedy

By using a phase correction algorithm or by widening the receive bandwidth.

 

Nyquist ghost artifact (diffusion image)

by radiographer nara


(한국어)

 

이피아이 기법의 나이키스트 유령 인공물?

데이터를 샘플링 할 때는 주파수폭을 1Hz 당 최소 2배의 신호를 주어야 그 주파수를 알 수 있다. 그러나 EPI 펄스 시퀀스의 경우 지그재그로 케이스페이스를 채워가기 때문에 각각의 위상부호화 단계에 대하여 에코가 중심점에서 약간씩 이동하게 된다. 이로 인해 나타나는 인공물을 나이퀘스트 유령 인공물이라고 하며, 위상방향으로 영상의 왜곡이 발생된다. FOV에 n개의 픽셀이 있다면, 이 인공물은 정확한 위치에 위치한 메인 이미지에 대해 n/2 픽셀만큼 이동 된 것으로 나타나며 3개로 나타난다.

 

방지대책

위상 수정 알고리즘을 사용하거나 수신대역폭을 넓게 해서 검사를 진행한다.

 