방사선사나라 Radiographer Nara

[MRI] (영/한) SENSE (Sensitivity - Encoded) / 센스기법 본문

자기공명영상 (MRI)

[MRI] (영/한) SENSE (Sensitivity - Encoded) / 센스기법

SEONARA 2020. 5. 2. 22:34
반응형

(영어/영문/English)

 

SENSE (=ASSET) ?

SENSE uses phase array coil (= synerge coil, = multi coil) to replace sequential gradient switching with parallel sensitivity encoding by a sensitive receiver.

The SENSE method acquires data by skipping the phase coding step like rFOV. In this case, aliasing artifacts are generated, and the SENSE technique acquires an image by separating the overlapped signal using a phase array coil.

Since the sensitivity of the coil is measured in advance, it has the advantage of obtaining an image without wrap around artifact in a short time while separating the signals at two points with a simple calculation, but the SNR is reduced because the number of phase edcoding is reduced.

 

A phase array coil is essential to obtain images in the SENSE method, and a reference scan is essential to measure sensitivity between the coil and the subject.

 

the signal A and B is overlapped.

 

seperating the signal A and B by using phase array coil and reference scan

 

by radiographer nara


(국어/국문/Korean)

 

SENSE (=ASSET) ?

SENSE는 phase array coil (=synergy coil, = multi coil)을 이용하여 감도를 가진 수신기에 의해 연속적인 경사 스위칭을 평행 감도 인코딩으로 대체하는 것이다.

센스 방식은 rFOV처럼 위상부호화 스텝을 건너뛰면서 데이터를 획득한다. 이렇게 되면 aliasing artifact가 발생하게 되는데, 센스 기법은 이 겹친 신호를 phase array coil를 이용하여 분리하여 영상을 획득한다.

미리 코일의 감도를 측정해 놓기 때문에 간단한 계산으로 두 지점의 신호를 분리하여 wrap around artifact가 없는 영상을 단축된 시간에 얻을 수 있는 장점이 있지만 위상부호화의 수가 줄어들기 때문에 SNR이 감소한다.

 

센스 방식으로 영상을 얻기 위해서는 phase array coil이 반드시 필요하고, 코일과 피사체간의 감도 측정을 위해 reference scan이 꼭 필요하다.

 

- 방사선사나라

반응형
Comments